OMA-300-CM, a process color monitor by high resolution diode array
spectroscopy. Providing tristimulus analysis using the visible spectrum from 340 nm to 720
nm. The results of the analysis can be displayed on screen and sent to the analog outputs,
or further processed to provide a single number such as the APHA value computed in
accordance with ASTM method D5386-93. This method converts CIE tristimulus color measurement of light
transmitted through near-clear liquid samples to equivalent visual color rating (Hazen or APHA). Process monitoring in the low 0-30 APHA range is feasible due to the multi
wavelength nature of the measurements. Also offering higher reproducibility and
accuracy in color measurements conforming with the following methods: Saybolt (D-156), Platinum
Cobalt (D-1209) and Dyed Aviation Gasoline (D-2392).
Principle of
operation
A diode array spectrophotometer measures
the total transmitted light through the flow cell, the measurements are converted, by the
analyzers CPU, to CIE X Y Z tristimulus color. These values are converted, if
required, to the yellowness index or any other color standard.
Verification
Verification of analyzer’s performance is carried out by the
manufacturer prior to shipment. The absorbance values of a solution of platinum
– cobalt must fall within a limit as specified by the ASTM method. The user will
use a filter to simulate these measurement in the field and to confirm
analyzer’s performance.
The OMA-300-CM process analyzer is offered
either as a general purpose or purged (explosion-proof) to comply with hazardous area
classification. All measurements are continuous. The analyzer utilizes fiber
optics to transmit light to and from the flow cell. Industry standard 4-20mA output
signals are available and easily re-scaled. User friendly interface, no
calibration required.
|