NOx and O2 on the inlet to an SCR. The gas stream contains ammonia, nitrogen oxides and in some cases sulfur dioxide . The NOx analyzer is a UV diode array process analyzer. Measuring the complete spectrum of the stream continuously and instantaneously.
The full
spectrum advantage
The system uses an aspirator
driven extractive sampling system with a heated umbilical sample line.
The sample flows through the absorption cell and the oxygen sensor in
series. Long sample lines are avoided, minimizing the normal
sample losses associated with sample adhesion to interior tubing walls.
|
|
| UV Absorbance spectra | |
![]() |
![]() |
![]() |
![]() |
![]() |
|
|
Specifications Range: 0 100 ppm NOx Alarm Contacts: Two (2) Range: 0 - 25% Oxygen Accuracy: + 2% of measured value, or + 0.1% O2, whichever is greater Repeatability: + 1% of reading, or 0.1% absolute, whichever is greater Response Time: Less than 4 seconds to 90% step change Signal Outputs 4-20mADC, isolated for Raw O2 4-20 mADC, isolated, for Raw NOx 4-20 mADC, isolated, for NOx, corrected to 15% O2 Local Area Network, RS-232, Power: 115VAC, 50/60 Hz Process: SCR Inlet Process Temp.: 800 ° F Process Pressure: supply: 29" of Water Column Return: Ambient Air |
|